Issue 1
2-2 Prepare the instrument K0449 - [EN] English
2.5 The battery
WARNING
This instrument uses a Lithium-Polymer (Li-Polymer)
battery pack. To prevent an explosion or fire, do not
short circuit, do not disassemble, keep it safe from
damage. For operating conditions, see Table 10-1.
To prevent an explosion or fire, use only the GE
specified battery, power supply and battery charger.
To prevent battery leakage or heat generation, only use
the battery charger and power supply in the
temperature range 0 to 40°C (32 to 104°F). For operating
conditions, see Table 10-1.
For a full battery specification, refer to Table 10-1.
2.5.1 Battery condition
When you first get your DPI 620 calibrator, the battery has
approximately 60% of full charge. You can use the battery
immediately but, for maximum operating time, we recommend
you give it a full charge (Section 2.5.3).
Charge indications When you set the power on, the battery symbol at the top of the
display shows the charge condition in 10% increments.
To get an accurate indication (1% increments), use the
Configuration menu; see Section 2.10.2 (Procedure to see the
instrument status).
2.5.2 Install the battery
b
2
a
1
Step Procedure
1. When the power is off, loosen the five screws
(a) and remove the cover (b).
If necessary, turn the instrument over and let
the discharged battery drop into your hand.
If the battery does not hold a charge, discard it
safely. Obey all the local health and safety
procedures.
2. Install the new battery correctly until it is flat in
the compartment.
3. Re-attach the cover.