[EN] English - K0449 Prepare the instrument 2-9
Issue 1
2.10.3 Procedures to make Task selections
When you first use the calibrator, there are default
measure and source functions on the display:
CH1 settings: RTD source, RTD type is PT100, scale is °C;
Automation is Nudge; see Chapter 3.
CH2 settings: Current measure, see Chapter 3.
Procedure Overview Use the Ta sk menu to complete these procedures:
Set the calibrator functions you want to use on the display:
item ; see Section 2.10.4.
a. CH1: Channel 1 electrical function: You can choose one
option:
b. CH2: Channel 2 electrical function: You can choose one
option:
c. * Pressure function (P1): for the PV 62x pressure
stations, refer to user manual - K0457; for the MC 620
module carrier, see Chapter 4.
d. * Pressure function (P2): the MC 620 module carrier can
use P1 and/or P2; see Chapter 4.
e. * IDOS function: Universal pressure module; see
Chapter 3.
f. HART function: HART device communications; see
Chapter 8
If applicable, change the Units for the function: item ;
see Section 2.10.5
CH1:
source
CH2:
measure
TAP
2
1
C
B
A
None Resistance
Current RTD
Voltage CJ
Millivolts AC Volts
TC (°C/°F) AC mV
TC (mV) AC Probe
Frequency Observed*
None Voltage
Current Millivolts
Current (24V) Observed*
A
Note: * Observed (Only
available as a source
option). Use this option to
make a manual record of
the readings on another
instrument; see
Section 2.11.3