[EN] English - K0449 Prepare the instrument 2-3
Issue 1
2.5.3 Charge the battery
You can charge the battery in the instrument (Section 2.6) or
you can use the optional external battery charger; see
Section 1.4 (Accessories).
2.5.4 Charge times
Note: You can use the instrument while the DC charger is
charging the battery but this increases the charge time.
2.5.5 Operating time
These are typical operating times for a new, fully charged
Li-polymer battery pack with these settings:
Backlight Intensity set to 80% (Default: 80%)
Backlight Timeout set to 2 hours (Default: 2 minutes)
Power save options To get the best battery duration, set a low value for the
Backlight Intensity (40%) and a short Timeout; see
Section 2.10.1 (Procedure to set the basic operations:)
2.6 The DC power
supply
WARNING
To prevent an explosion or fire, use only the GE
specified battery, power supply and battery charger.
To prevent battery leakage or heat generation, only use
the battery charger and power supply in the
temperature range 0 to 40°C (32 to 104°F). For operating
conditions, see Table 10-1.
This unit supplies power to the instrument and charges the
battery (if it is installed). All the instrument functions are
available with or without the battery.
Charge method Charge time (to full capacity)
DC power supply
| 6.5 hours
External battery charger
| 6.5 hours
USB mini Type B connector
| 13 hours (with a 500 mA
supply)
Operation Battery duration
Continuous operation (measure) > 10 hours
Continuous operation (measure
and source with loop power on)
> 6 hours