[EN] English - K0449 HART
®
device operations 8-9
Issue 1
8.5.2 HART menu - Information (Advanced, Clone)
Example procedure: Clone
The Clone function lets you copy the parameters from one field
device to another of the same type and range.
1. Attach the first device with the necessary parameters and
start communications (Section 8.5).
2. Go to the Clone menu option and tap on Copy. It copies
these parameters:
3. Go back to the Home display and attach the other field
device.
4. Start communications with the new device (Section 8.5).
5. Go to the Clone menu option and tap on Paste.
To complete the Clone operation from one device to another,
make sure the DPI 620 calibrator stays on for the whole
procedure.
1
See
Section 8.5.1
D
E
F
A
B
C
Tag Damping value
PV unit Transfer function
LRV (lower range value)
URV (upper range value)

Table of Contents