Issue 1
3-4 Electrical and IDOS operations K0449 - [EN] English
1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections and continue with the
measure operation.
3.2.5 Example procedure: Measure AC voltage (CH1) with the AC probe
WARNING
To prevent electrical shocks, use only the GE specified
AC probe (Part: IO620-AC) to measure AC voltages that
are more than 20 Vrms (maximum: 300 Vrms). Attach it
to the specified connections only.
This example shows Channel 1 (CH1) set up to measure an
AC voltage with the AC probe (maximum: 300 Vrms).
1. Set the applicable software options for the AC Probe; see
Section 3.2.1 (Procedure overview).
2. Complete the electrical connections.
Red: V/Hz connector; Black: COM connector
Then continue with the measure operation.
2
Measure AC millivolts on
channel 1 (CH1)
Range: 0 to 2000 mV(rms)
1
B
Measure AC volts on
channel 1 (CH1)
Range: 0 to 20 V(rms)
A
1
Maximum: 0 to 20 Vrms
AC
21
AC
Maximum: 300 Vrms