[EN] English - K0449 General specification 10-1
Chapter 10: General specification
10.1 Introduction
For a full specification of the DPI 620 calibrator and its related
accessories (MC 620 carrier, PM 620 module, and
PV 62x pressure stations) refer to the datasheet supplied on
the CD (CD: P/N UD-0002).
Table 10-1: General specification
Display LCD: Colour display with touch-screen
Operating temperature -10 to 50°C (14 to 122°F)
Storage
temperature
-20 to 70°C (-4 to 158°F)
Ingress Protection IP65 (DPI 620 calibrator only)
Humidity 0 to 90% relative humidity (RH) non-condensing
Shock/Vibration Def Stan 66-31, 8.4 cat III
EMC Electromagnetic compatibility: BS EN 61326-1:2006
Electrical safety Electrical - BS EN 61010:2001
Pressure safety Pressure Equipment Directive - Class: Sound Engineering Practice
(SEP)
Approved CE Marked
Battery power Lithium-Polymer battery (GE Part number: 191-356)
Capacity: 5040 mAh (minimum), 5280 mAh (typical); Nominal voltage:
3.7 V.
Charge temperature: 0 to 40°C (32 to 104°F)
When the instrument senses the temperature is outside this range, it
stops charging.
Discharge temperature: -10 to 50°C (14 to 122°F).
Charge/discharge cycles: > 500 > 70% capacity

Table of Contents