[EN] English - K0449 Electrical and IDOS operations 3-1
Chapter 3: Electrical and IDOS
operations
3.1 Introduction
This section gives examples of how to connect and use the
instrument for these operations:
to measure and source electrical values
to measure the pressure with an IDOS module
Before you start:
Read and understand the “Safety” section.
Do not use a damaged instrument.
Note: Use only original parts supplied by the manufacturer.
3.2 Measure and
source operations
When you first use the calibrator, there are default
measure and source functions on the display; see
Section 2.10.3.
3.2.1 Procedure overview
To use the calibrator, complete these procedures:
Set the calibrator functions you want to use on the display:
item ; see Section 2.10.3. This includes:
a. CH1: Channel 1 electrical function (measure or source)
b. CH2: Channel 2 electrical function (measure or source)
c. Pressure function (P1 and/or P2): for the MC 620 module
carrier, see Chapter 4; for the PV 62x pressure stations
(P1 only), refer to user manual - K0457
d. IDOS function: to measure pressure with an IDOS
universal pressure module (UPM), see Section 3.3
e. other functions (maximum: 6 functions)
If necessary, change the Units for the function: item ;
see Section 2.10.5
If necessary, set a Utility for the function: item ; see
Section 2.10.6
a. Max/Min/Avg
b. Switch Test
c. Leak Test (pressure options only)
C
B
A
TAP
2
1
A
B
C