[EN] English - K0449 Datalog operation 6-3
Issue 1
6.3 Data Logging
To data log:
To log measurements made by the instrument, set the required
tasks in Task Settings.
Select Configuration and select Data Logging.
Select Filename and enter a name using the three-page screen
alpha/numeric key-pad.
Select Trigger and either a time (Periodic) or key press
(play/pause).
Select start to set the data logging.
The display changes to the task screen with data logging
flashing at the top.
If necessary, press play/pause to increment the data logging.
At the end of data logging press the cancel key.
The instrument stores the recorded data.
6.4 Data handling
Using the data logging menu, a data file can be replayed,
stored in a memory device or transferred to an external device;
see Section 6.2.
The data file can be imported into a number of spreadsheet
programs for analysis and graphic presentation.
Comma separation applies to this data.
Comma separated file Data imported into a spreadsheet