[EN] English - K0449 Electrical and IDOS operations 3-3
Issue 1
1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections and continue with the
measure or source operation.
3. Source only (Automation): Set the applicable output value;
see Section 2.11.2.
3.2.3 Example procedure: Measure DC voltage
These examples (A and B) show Channel 1 (CH1) set up to
measure a DC voltage.
Note: If you use the Channel 2 (CH2) connectors, you can use
Channel 2 (CH2) to measure these ranges.
1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections and continue with the
measure operation.
3.2.4 Example procedure: Measure AC voltage (CH1), 0 to 20 Vrms only
WARNING
To prevent electrical shocks, use only the GE specified
AC probe (Part: IO620-AC) to measure AC voltages that
are more than 20 Vrms (maximum: 300 Vrms). Attach it
to the specified connections only. See Section 3.2.5.
These examples (A and B) show Channel 1 (CH1) set up to
measure an AC voltage (0 to 20Vrms only).
2
Measure DC millivolts on
channel 1 (CH1)
Range: ±2000 mV (DC)
1
B
Measure DC volts on
channel 1 (CH1)
Range: ±30 V (DC)
A
1