Issue 1
3-6 Electrical and IDOS operations K0449 - [EN] English
1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections and continue with the
measure or source operation.
3. Source only (Automation): Set the applicable output value;
see Section 2.11.2.
3.2.8 Example procedure: Measure or source frequency signals
These examples (A and B) show Channel 1 (CH1) set up to
measure or source a frequency. This includes Hz, kHz and
counts (cpm or cph).
The selection of units sets the available range, for example:
Example A 1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections.
3. If necessary change the Trigger Level (Settings) and
continue with the measure operation.
Values to set:
Mode (Automatic/Manual);
Manual Level (trigger level value)
Example B 1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
Hz = 0 to 1000 Hz kHz = 0 to 50 kHz
21
Measure frequency on
channel 1 (CH1)
Range: 0 to 50 kHz
Trigger level: 2.5 V
A
3
Source frequency on
channel 1 (CH1)
Range: 0 to 50 kHz
Waveform: Triangle
Amplitude: 12.0 V
Automation: Nudge
(Section 2.11.2)
2B
4
1
3