Issue 1
3-8 Electrical and IDOS operations K0449 - [EN] English
Values to set:
RTD Type (Set the applicable RTD); see Table A 1 (front
cover) for the available options.
Example B 1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections.
3. If necessary change the Source Settings and continue with
source operation.
Values to set:
RTD Type (Set the applicable RTD); see Table A 1 (front
cover) for the available options.
4. Automation: Set the applicable output value; see
Section 2.11.2.
3.2.10 Example procedure: Measure or simulate a thermocouple (or TC mV)
These examples (A and B) show Channel 1 (CH1) set up to
measure or simulate a thermocouple temperature.
Note: To measure or simulate TC millivolts, set the TC mV
function
Example A 1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections.
21
Measure a TC temperature
on channel 1 (CH1)
Range: 1372 °C
TC Type: K Type
Cold Junction: 32.66
A
3
Simulate a TC temperature
on channel 1 (CH1)
Range: 1372 °C
TC Type: K Type
Cold Junction: 32.66
Automation: Nudge
(Section 2.11.2)
2B
4
1
3