EasyManua.ls Logo

GE Druck PM 620

GE Druck PM 620
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Issue 1
2-16 Prepare the instrument K0449 - [EN] English
Percent Step
Defined Step
Ramp
2.11.3 Set the Observed settings (source )
The Observed function is available as a Source on five of the
task selections: CH1, CH2, P1, P2, and IDOS. To set the
Observed function, see Section 2.10.4.
1
TAP
Description
Process: Percent Step;
Values to set: Low, High, Step Size (%FS), Dwell,
Auto Repeat (On/Off)
Use these buttons to change the value manually by the
specified Step Size.
Use the start and stop buttons to change the value
automatically by the specified Step Size.
Dwell sets the period between each change.
1
TAP
Description
Process: Defined Step;
Values to set: Low, High, Step Size, Dwell, Auto Repeat (On/Off)
Use these buttons to change the value manually by the
specified Step Size.
Use the start and stop buttons to change the value
automatically by the specified Step Size.
Dwell sets the period between each change.
1
TAP
Description
Process: Ramp;
Values to set: Low, High, Travel, Dwell, Auto Repeat (On/Off)
Use the start and stop buttons for this process. Travel
sets the duration of the Ramp.
Dwell sets the period between each change in direction.