[EN] English - K0449 Prepare the instrument 2-17
Issue 1
You can use the Observed function to make a manual record of
the readings you can see on another instrument. For example,
you can make a permanent record of the readings with
Data Logging (Chapter 6).
Settings These settings are available:
2.12 The Advanced
menu options
The Advanced menu lets you set or use these options:
2.12.1 Advanced: Calibration options
There are two calibration menus you can use:
Menu to calibrate the
DPI 620
1. Enter the calibration PIN: 4321
2. Tap the “Accept” button. You can then continue with these
operations:
• Calibrate the different measure and source channels on the
DPI 620 calibrator.
Contact us at: www.gesensinginspection.com
• Change the calibration PIN. You must confirm the new PIN.
1
TAP
Description
Values to set:
Units (enter a name for the units)
Resolution (Set one of the 7 options: 1.0 to 0.000001)
Use this button (Setpoint) to set the values you see on the
other instrument.
• HART resistor • Calibration
• Standards • Setup Files
2
1
Set the applicable temperature scale for all
temperature operations; see
Chapter 3
Options: IPTS 68 or ITS 90.
Set On or Off; see
Chapter 8
See Section 2.12.1
See Section 2.12.2
3
TAP
1
TAP
2