Issue 1
2-18 Prepare the instrument K0449 - [EN] English
• Set the next calibration date and/or set the Notification
option. If Notification is On and your calibration is overdue,
you get a “Calibration due ... ” message at the end of the
power on sequence.
Menu to upgrade the
DPI 620 software and
firmware
To use most up to date software and firmware on your DPI 620
calibrator, visit our website:
www.gesensinginspection.com
Follow the website instructions to download the files onto a
USB flash memory drive. You can then use this menu to
upgrade your calibrator.
1. Enter the calibration PIN: 5487
2. Tap the “Accept” button. You can then continue with one of
these operations:
• Upgrade the software. To complete this operation, put the
USB flash memory drive in the USB type A connector
(Figure 1-3) and follow the on-screen instructions.
• Upgrade the sensor firmware (PM 620). To complete this
operation, put the USB flash memory drive in the USB type
A connector (Figure 1-3) and follow the on-screen
instructions.
• Upgrade the HART Firmware. To complete this operation,
put the USB flash memory drive in the USB type A
connector (Figure 1-3) and follow the on-screen instructions.
Note: If you make a mistake and there are no files to upload,
follow the on-screen instructions and complete the procedure.
When an upgrade completes normally, the initial operation of
the touch screen is slower (a period of approximately 30
seconds).
To make sure the upgrade completed correctly, use the Status
menu (Configuration > Status > Software Build); see
Section 2.10.2.
1
TAP
2