Issue 1
7-2 Documenting functions K0449 - [EN] English
Input and Reference
options
Channel type : Input or Reference
Scaling (Input only): Values for Reference High and Low
and Input High and Low. This sets the scale for the Analysis
function.
Error Type (Input only): % Span or % Rdg (Reading)
Linearity (Input only): Linear or Square Root
Tolerance (Input only): an applicable value for the
pass/fail calculation
When the Input and Reference options are set, start the
analysis.
7.3 Run a procedure
The Intecal calibration Procedure contains all the values to
calibrate a device under test (test points, ramp time). You can
use the same calibration procedure for all the applicable
devices under test.
To use the Run Procedure function, you need these items:
a copy of the Intecal Calibration Software. This is available
from our website at: www.gesensinginspection.com
There is a 30 day free trial.
IO620-USB-RS232 (Intecal version 5 only). An adaptor to
connect the DPI 620 calibrator to the RS232 interface on
your computer.
a DPI 620 calibrator device driver: available as a download
from our website.
A
B
C
D
E
Set applicable values. When the Analysis
work is complete, tap on exit.
Start the
Analysis
9
10
Continued
from step 8
in specified
tolerance limit
not in specified
tolerance limit