Chapter 7: Documenting functions
7.1 Introduction
This section gives examples of the documenting functions
available with the DPI 620 calibrator. There are two options:
Analysis (Section 7.2): This function lets you compare the
data from two Channels on the DPI 620 calibrator: the device
under test (DUT) and a reference instrument.
It calculates the % span or % reading error for the device and
then gives a pass or fail indication.
Run Procedure (Section 7.3): This uses the Intecal
Calibration software (Optional item) to download a calibration
procedure.
Before you start:
• Read and understand the “Safety” section.
• Do not use a damaged instrument.
Note: Use only original parts supplied by the manufacturer.
7.2 Analysis
To use the Analysis function, set the functions for the two
Channels you want to compare (Section 2.10.4). Then use
these procedures: