5-84 Section 5-11 - Common Service Desktop (R3 only)
Clean Userdefs
Clean Userdefs provides a way to delete user-defined data when user presets may be inhibiting proper
operation. User-defined data returns to the user-defined settings to the factory default values. Clean
Userdefs provides information to backup the current settings if you want to use them later. Disruptive
mode must be On to clean user-defined data. If Disruptive mode is Off, a warning message displays
asking to turn on Disruptive mode.
Before cleaning user-defined data, the LOGIQ P9/P7 application software automatically shuts down.
Once the operation is complete, the system reboots.
Clean Userdefs provides a way to selectively delete the following user-defined data:
Connectivity – Deletes this file: %TEST_ROOT%\resources\userdefs\connectivity.res
Imaging – Deletes this folder: %TEST_ROOT%\resources\userdefs\ImagingPresets\
Measurement – Delete this folder: %TEST_ROOT%\resources\userdefs\EchoMeasure\
Annotation/Bodymark – Deletes the folder: %TEST_ROOT%\resources\userdefs\EchoAnnotation\
Scan Assistant – Deletes the folder: %TEST_ROOT%\resources\userdefs\EchoExamProtocol
The information on Clean Userdefs is available to Class C (Service Advanced) and Class M licenses.
To access Clean Userdefs, select Utility (second page) > Service > Utilities > Clean Userdefs.
Figure 5-57 Clean Userdefs
This table shows all the elements available on Clean Userdefs with descriptions.
Table 5-36 Clean Userdefs
Element DESCRIPTION
All
When Disruptive mode is ON, deletes all of the user-defined data and returns the
LOGIQ P9/P7 to factory default settings.
Selective Delete When Disruptive mode is ON, deletes the selected user-defined data.
Connectivity
When Selective Delete is selected, deletes this file:
%TEST_ROOT%\resources\userdefs\connectivity.res
Imaging
When Selective Delete is selected, deletes this folder:
%TEST_ROOT%\resources\userdefs\ImagingPresets\
Measurement
When Selective Delete is selected, delete this folder:
%TEST_ROOT%\resources\userdefs\EchoMeasure\
Annotation/Body Patterns
When Selective Delete is selected, deletes this folder:
%TEST_ROOT%\resources\userdefs\EchoAnnotation\
Scan Assistant
When Selective Delete is selected, deletes this folder:
%TEST_ROOT%\resources\userdefs\EchoExamProtocol
Clean Userdefs Restarts the application software with a fresh database.

Table of Contents

Other manuals for GE LOGIQ P7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P7 and is the answer not in the manual?

GE LOGIQ P7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P7
CategoryMedical Equipment
LanguageEnglish

Related product manuals