GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5-85
To clean user-defined data:
1.) If you want to use the current settings later, back them up now.
2.) Navigate to Utility (second page) > Service > Utilities > Disruptive Mode Utility and set
Disruptive mode to On.
3.) Navigate to Utility (second page) > Service > Utilities > Clean Userdefs.
4.) Select the data you want to set to the default settings.
5.) Click Clean Userdefs.
For more information, see: