Connecting a vaporizer during Run mode
Insert the vaporizer into the vaporizer slot and
push into place until a click is heard.
The vaporizer check cannot be performed
during Run mode. A 'Vaporizer check
bypassed' message will be displayed on the
screen, and an entry is made in the event log.
When connecting a vaporizer during Run
mode, only connect vaporizers that previously
have passed System checkout/Vaporizer
check.
WARNING! Bypassing the Vaporizer check
removes the possibility to detect leakage
between the vaporizer and main system.
Agent may thus leak out into the room
undetected.
136
FLOW-i 4.2, User's Manual