Upper and lower alarm limits
To change a parameter setting, press its touch
pad and use the rotary knob to increase or
decrease the setting. A parameter can also be
selected by moving through the values using
the rotary knob, and pressing the rotary knob
when the blue box highlights the desired
parameter touch pad.
New parameter settings are confirmed and
applied to the system when the Accept touch
pad is pressed.
Alarm sound level
The alarm sound level can be set from 20%
(approx. 55 dB) to 100% of the maximum
(approx. 85 dB).
WARNING! Ambient sounds levels may
impede the recognition of auditory alarm
signals. Make sure the alarm sound level is
set sufficiently high.
Enabling/Disabling certain alarms
Some alarms can be disabled by selecting the
associated parameter and turning the rotary
knob counter-clockwise (lower limit settings)
or clockwise (upper limit settings) until the
Alarm OFF symbol appears. These include:
FiO
2
: High
EtO
2
: High
EtO
2
: Low
EtCO
2
: Low
FiAA: Low
EtAA: Low
In manual (MAN) mode, the following
additional alarms can be disabled:
Expiratory Minute Volume: Low
Expiratory Minute Volume: High
Respiratory Rate: Low
Respiratory Rate: High
Apnea
Alarm autoset
In controlled ventilation modes, the upper and
lower limits of the ventilation metric alarms
can be automatically set by the system during
Run mode. Gas metric alarms cannot be
automatically set this way.
The autoset limits are derived from the
measured patient values displayed on the
screen. A percentage of the displayed value,
or a fixed number, is either added to, or
subtracted from the displayed value to give
the upper and lower alarm limits. See the
following tables for specific values.
188
FLOW-i 4.2, User's Manual
| 9
| Alarms and patient safety |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals