9.2.1 Ventilation alarm limit settings (Adult)
Alarm type
AutosetStepDefault
Upper
limit
Lower
limit
Setting
unit
Alarm
type
Alarm
+10 cmH
2
O
(min 35)
1 cmH
2
O40 cmH
2
O120 cmH
2
O10 cmH
2
OcmH
2
O
5
UpperPpeak
-50 %0.5 l/min2 l/min
OFF (MAN)
40 l/min0.5 l/min
OFF (MAN)
l/minLowerMVe
+50 %0.5 l/min20 l/min
OFF (MAN)
60 l/min
OFF (MAN)
0.5 l/minl/minUpperMVe
-40 %1 b/min4 b/min
OFF (MAN)
140 b/min1 b/min
OFF (MAN)
B/minLowerRR
+40 %1 b/min20 b/min
OFF (MAN)
140 b/min
OFF (MAN)
1 b/minB/minUpperRR
-3 cmH
2
O1 cmH
2
O2 cmH
2
O47 cmH
2
O0 cmH
2
OcmH
2
O
6
LowerPEEP
(AUTO)
+5 cmH
2
O1 cmH
2
O10 cmH
2
O55 cmH
2
O0 cmH
2
OcmH
2
O
6
UpperPEEP
(AUTO)
5. Available units include cmH
2
O and mbar
FLOW-i 4.2, User's Manual
189
| Alarms and patient safety |
9 |