9.2.2 Ventilation alarm limit settings (Infant)
AutosetStepDefault
Upper
limit
Lower
limit
Setting
unit
Alarm
type
Alarm
+10 cmH
2
O
(min 35)
1 cmH
2
O25 cmH
2
O90 cmH
2
O10 cmH
2
OcmH
2
O
6
UpperPpeak
-50 %0.01-0.1
l/min
0.8 l/min
OFF (MAN)
20 l/min0.01 l/min
OFF (MAN)
l/minLowerMVe
+50 %0.01-0.1
l/min
3 l/min
OFF (MAN)
30 l/min
OFF (MAN)
0.01 l/minl/minUpperMVe
-40 %1 b/min15 b/min
OFF (MAN)
140 b/min
OFF (MAN)
1 b/min
OFF (MAN)
B/minLowerRR
+40 %1 b/min50 b/min
OFF (MAN)
140 b/min1 b/minB/minUpperRR
-3 cmH
2
O1 cmH
2
O1 cmH
2
O47 cmH
2
O0 cmH
2
OcmH
2
O
6
LowerPEEP
(AUTO)
+5 cmH
2
O1 cmH
2
O7 cmH
2
O55 cmH
2
O0 cmH
2
OcmH
2
O
6
UpperPEEP
(AUTO)
6. Available units include cmH
2
O and mbar
190
FLOW-i 4.2, User's Manual
| 9
| Alarms and patient safety |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals