9.2.3 Gas alarm limit settings
AutosetStepDefault
Upper
limit
Lower
limit
Setting
unit
6
Alarm
type
Alarm
---0.1%1%10%0.1%[gas]%UpperFiCO
2
---0.1%4%
OFF (MAN)
9.9%OFF, 0.1%[gas]%LowerEtCO
2
---0.1%6.5%10%0.1%[gas]%UpperEtCO
2
---1%25 %99%18%[gas]%LowerFiO
2
---1%OFF99%, OFF23%[gas]%UpperFiO
2
---1%OFF99%OFF, 10%[gas]%LowerEtO
2
---1%OFF99%, OFF13%[gas]%UpperEtO
2
---0.1%OFF5.0% (ISO)
8.0% (SEV)
18.0%(DES)
or OFF
OFF, 0.1%[gas]%LowerFiAA
---0.1%4.0% (ISO)
8.0% (SEV)
16.0%(DES)
5.0% (ISO)
8.0% (SEV)
18.0%(DES)
or OFF
0.1%[gas]%Upper
7
FiAA
---0.1%OFF4.0% (ISO)
6.0% (SEV)
12.0%(DES)
OFF, 0.1%[gas]%LowerEtAA
---0.1%2.3% (ISO)
4.0% (SEV)
12.0%(DES)
5.0% (ISO)
8.0% (SEV)
18.0%(DES)
or OFF
0.1%[gas]%UpperEtAA
6. Available units include [gas]%, kPa (partial pressure) and mmHg (partial pressure)
7. Alarm is also activated if the MAC is above a certain value for a defined period of time
FLOW-i 4.2, User's Manual
191
| Alarms and patient safety |
9 |