9.3.2 Pre-muting alarms
Alarms can be pre-muted by depressing the
membrane button for more than two seconds.
When pressed, a timer counting down from
two minutes is displayed in the 'Audio pause
and Countdown timer' area.
2:00
Press again for less than two seconds and a
new two minute pre-mute period is started.
Pressing for more than two seconds during
the countdown period will deactivate the
pre-mute function.
Different symbols are used to differentiate
between the mute alarm timer and the
pre-mute alarm timer.
9.3.3 Mute until further notice (Audio OFF)
Some alarms, e.g. 'FiO2: High' and 'Apnea
60/120 s', can be muted until further notice,
or until the alarm condition resolves.
Pressing the audio off button to silence any
of these alarms will produce a dialog window:
'Do you want to turn <applicatory alarm>
audio OFF?'. Selecting 'Yes' will mute the
alarm until further notice. Selecting 'No' will
mute the alarm for two minutes.
FLOW-i 4.2, User's Manual
193
| Alarms and patient safety |
9 |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals