DescriptionAlarm
Activated when inspiratory and expiratory pressures
fail to meet preset requirements
Check breathing circuit
Inspiration end pressure - Inspiration start pressure
< 1 cmH
2
O for three consecutive breaths
Excessive leakage
Measured inspiratory O
2
concentration is below the
set Lower alarm limit.
FiO
2
: Low
Inspiratory N
2
O gas supply >80%FiN
2
O: High
Inspiratory agent concentration exceeds set upper
limit.
FiAA: High
Expiratory agent concentration exceeds set upper
limit.
EtAA: High
Less than 18 minutes left of battery operation.Limited battery capacity
Less than 3 minutes left of battery operation.No battery capacity
Connection error; unable to estimate battery time.Battery operation - battery capacity unknown
No cassette is connectedPatient cassette disconnected
FLOW-i 4.2, User's Manual
197
| Alarms and patient safety |