9.8 Gas supply
9.8.1 Loss of central gas pressure
If the O
2
, Air or N
2
O gas supply pressure falls
below approx. 250 kPa (2.5 bar, 26 Psi), that
particular gas supply is switched off and the
'Gas supply: Low' alarm is triggered. Gas
delivery subsequently switches according to
the table below. Supply of the missing gas
automatically resumes when the supply
pressure of that gas rises above approx. 280
kPa (2.8 bar, 40 Psi). Following this, the
system reverts to delivering gas according to
the configurations prior to the loss of gas
supply.
Gas deliveryMissing gasGas mix
O
2
/AirN
2
OO
2
/Air
O
2
N
2
OO
2
/N
2
O
O
2
AirO
2
/Air
O
2
/N
2
OAirO
2
/N
2
O
AirO
2
O
2
/Air
AirO
2
O
2
/N
2
O
---AIR and O
2
O
2
/Air
WARNINGS!
• If the O
2
supply is lost, delivery of
anesthetic agent will be stopped.
• If the Air supply is lost, the patient
suction unit (option) will not function.
9.8.2 Backup gas supply (option)
The backup gas supply acts as a reserve
system should the central gas supply become
unavailable.
+-
+
-
O
2
Air
O
2
Air
B
A
The C20 and C30 systems can be equipped
with a backup gas rack (A) configured to
deliver either O
2
/Air or O
2
/N
2
O and a backup
gas holder (B) delivering N
2
O.
204
FLOW-i 4.2, User's Manual
| 9
| Alarms and patient safety |