10.3 AFGO
AFGO ventilation is carried out using an
external breathing system connected to the
system's AFGO (Additional Fresh Gas Outlet)
connection.
This mode bypasses the circle system,
creating an open system with gas and agents
primarily moving in one direction. Re-use of
gas and CO
2
clearance depends on the
current fresh gas flow.
Refer to manufacturers description for
recommendations on minimal flow required
to carry out ventilation using any specific
external breathing system.
For a complete description on AFGO, refer to
Chapter 8 - AFGO.
CAUTION: Only external breathing circuits
equipped with a pressure relief valve or a
manual bag with an opened end shall be
connected to the AFGO outlet.
10.3.1 Important considerations
The following alarms are disabled during
AFGO ventilation:
Expiratory Minute Volume: High
Expiratory Minute Volume: Low
Apnea > 60 s / Apnea > 120 s
Continuous APL pressure
Refer to Chapter 10, page 202, for more
information on alarm functions in AFGO mode.
FLOW-i 4.2, User's Manual
211
| Ventilation modes |
10 |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals