Inspiration criteria
The initiation of inspiration is based on the
preset breathing frequency, or when the
patient triggers a breath. The duration of
Inspiration is based on the breathing frequency
and the I:E ratio, or the breathing frequency
and Ti.
Pause time
The pause time between inspiration and
expiration can be set from 0% to 30% of the
breath cycle time. Default setting is 10%.
10.6.2 Important considerations
WARNING! Volume control is not intended
for use on neonates because the Tidal
volume (V
T
) cannot be set < 20 ml.
Set upper pressure alarm limit. If this limit
is reached, inspiration will cease. Expiration
will begin and an alarm for high airway
pressure will be triggered.
Carefully monitor patient airway pressures.
If the patient has poor lung characteristics
there is a risk of high airway pressures.
Discrepancies in gas concentrations
between the patient’s inspiration and
expiration depend on the patient’s uptake.
The patient can trigger extra breaths
resulting in a higher delivered minute
volume than the set one.
218
FLOW-i 4.2, User's Manual
| 10
| Ventilation modes |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals