10.9.3 SIMV (Pressure Control) + Pressure Support
90%
31 2
The following parameters are set:
1. PC (Pressure Control level) above PEEP
(cmH
2
O)
2. SIMV rate (b/min)
3. PEEP (cmH
2
O)
4. I:E ratio / Insp. time
5. Inspiratory rise time (%/s)
6. Breath cycle time
7. Inspiratory Cycle-off (%)
8. Trigg. Flow / Trigg. Pressure
9. PS (Pressure Support level) above PEEP
(cmH
2
O)
10. Gas mix (O
2
/Air or O
2
/N
2
O)
11. O
2
concentration (%)
12. Fresh gas flow (l/min)
13. Agent concentration (%)
SIMV (PC) properties
This combined control and pressure
support/spontaneous function allows for
preset mandatory breaths synchronized with
the patient's breathing.
1. The mandatory breath is defined by the
basic settings (mode of ventilation, breath
cycle time, respiratory pattern and
volumes/pressures).
2. The spontaneous/pressure supported
breaths are defined by the setting for
Pressure Support.
3. If there is no trigger attempt within a time
window equal to 90% of the set Breath
cycle time, a mandatory breath is
delivered. (The Breath cycle time is the
total time for one mandatory breath.)
228
FLOW-i 4.2, User's Manual
| 10
| Ventilation modes |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals