EasyManua.ls Logo

Getinge Maquet FLOW-i 4.2

Getinge Maquet FLOW-i 4.2
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Expiration criteria
Expiration begins:
When the inspiratory flow falls below a
preset fraction of the inspiratory peak flow
(Inspiratory cycle-off)
If the upper pressure limit is exceeded.
Maximum time for inspiration is exceeded.
Inspiration criteria
When the patient triggers a breath
according to the set trigger sensitivity.
Mandatory breath requirement is met during
the breath cycle time.
Triggered breaths are highlighted in the
waveform display area, see Chapter 5, page
103 for more information.
FLOW-i 4.2, User's Manual
229
| Ventilation modes |
10 |

Table of Contents

Related product manuals