3. Remove the patient cassette by lifting the
cassette upwards and out using the
intended grip located between the
inspiratory and expiratory connections.
4. Remove the caps and valve cages from
the inspiratory/expiratory valves (turn
counter-clockwise). Place the valve cages
containing the ceramic disks in a fine
mesh box and re-attach the caps.
240
FLOW-i 4.2, User's Manual