12.5 Wipe off/discard
Dismantle
Wipe off/discard
Rinse
DryDisinfectant
Rinse
Washer-
disinfector
Dry
Assemble
System checkout
Steam
autoclave
Wipe the exterior of the system, the patient
cassette compartment, the volume reflector
compartment and all removable parts with a
soft lint-free cloth moistened in soap (pH <10)
and water or a mild detergent-based
disinfectant.
In the case of more contaminated surfaces,
the following products have been approved
for use:
Alcohol (ethyl or isopropyl alcohol 70%)
Wipe the system free of soap and disinfectant
residues using a lint-free cloth moistened in
water only.
Do not pour or spray fluid into the patient
cassette compartment.
FLOW-i 4.2, User's Manual
243
| Routine cleaning and maintenance |
12 |