Discard disposable items according to local
and environmental regulations.
Regularly check that the fan filter looks clean
(i.e. black). If dusty, remove (snap off/snap on)
and rinse in water. Shake out all excess water
from the filter.
After removing the patient cassette and
volume reflector, avoid pouring or spraying
any liquid into the open compartments. Avoid
contact with electronics and electrical
connectors when wiping.
Check the Suction unit bacterial/viral filter after
each patient case for signs of moisture or fluid
and replace if necessary. Fluid in the filter will
compromise the functionality of the filter and
Suction unit.
Firmly pull the filter outwards and dispose it
according to hospital routines for
contaminated waste products.
Attach the new filter by fitting the unmarked
end into the back of the suction module and
press until a click is heard. Ensure the filter is
correctly secured in the locked position.
244
FLOW-i 4.2, User's Manual
| 12
| Routine cleaning and maintenance |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals