4. For optimal cleaning results, fit the
disassembled parts into the
washer-disinfector according to the
suggestion in the illustration below:
To further increase the washing efficiency,
plug the remaining water outlets to increase
the water flow through the patient cassette
and volume reflector.
The water pressure in the
washer-disinfector should not exceed 1.5
bar.
The water flow should never be less than
10 l/min.
The water must have free passage through
the cassette.
The patient cassette and volume reflector have
been tested to withstand 200 disinfection
cycles in a washer disinfector.
Proceed to section 11.9, page 253, for drying
instructions.
CAUTION: Water remaining in the patient
cassette and volume reflector after a
completed washing program may be very
hot.
250
FLOW-i 4.2, User's Manual
| 12
| Routine cleaning and maintenance |