12.10.1 Assembling volume reflector
1. Make sure the lever in the cassette
compartment securing the volume
reflector is correctly positioned.
2. Re-insert the volume reflector.
3. Plug in the volume reflector adaptor and
re-attach the safety valve if removed.
For devices with serial numbers below 5001,
the volume reflector adaptor has a steel wire
retaining bracket. For devices with serial
numbers above 5001, the volume reflector
adaptor is delivered without this retaining
bracket. The adaptor without the retaining
bracket can be used on all FLOW-i devices,
while the adaptor with the retaining bracket
may NOT be used for devices with serial
numbers above 5001. To avoid any risk of
confusion, it is a good idea to remove the
retaining bracket from the volume reflector
adaptor.
FLOW-i 4.2, User's Manual
257