13.11 Breathing circuits and accessories
Note that the table applies to the breathing circuits recommended by Maquet.
Breathing circuits and accessories
Compliance 10 mm breathing circuits: <0.4 ml/cmH
2
O
15 mm breathing circuits: <0.7 ml/cmH
2
O
22 mm breathing circuits: <1.4 ml/cmH
2
O
Internal volume 10 mm breathing circuits: 0.4 l
15 mm breathing circuits: 0.7 l
22 mm breathing circuits: 1.4 l
Flow resistance in each limb
including Y-piece
10 mm breathing circuits:
At 2.5 l/min: <3.0 cmH
2
O/l/s
At 15 l/min: <6.0 cmH
2
O/l/s
At 30 l/min: <10 cmH
2
O/l/s (30 l/min not applicable for intended
patient tidal volume range)
15 mm breathing circuits:
At 2.5 l/min: <1.5 cmH
2
O/l/s
At 15 l/min: <1.5 cmH
2
O/l/s
At 30 l/min: <2 cmH
2
O/l/s
22 mm breathing circuits:
At 2.5 l/min: <0.5 cmH
2
O/l/s
At 15 l/min: <0.7 cmH
2
O/l/s
At 30 l/min: <1.0 cmH
2
O/l/s
Flow resistance for angled
adapter
At 2.5 l/min: <0.2 cmH
2
O/l/s
At 15 l/min: <0.6 cmH
2
O/l/s
At 30 l/min: <1.0 cmH
2
O/l/s
Flow resistance for 22 mm joint
adapter
At 2.5 l/min: <0.2 cmH
2
O/l/s
At 15 l/min: <0.2 cmH
2
O/l/s
At 30 l/min: <0.2 cmH
2
O/l/s
FLOW-i 4.2, User's Manual
275
| Technical specifications |
13 |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals