13.12 Ventilator
Ventilator
Pneumatic powered Servo controlledType
Neonatal to AdultPatient range
• Manual/Bag
• AFGO (option)
• Volume Control (VC)
• Pressure Control (PC)
• Pressure Support (PS, option)
Ventilation modes
20 – 2000 ml, ±10% or 10 ml, whichever is greaterTidal volume range
(volume controlled modes)
20 - 350 ml, resolution 1 mlInfant range:
100 - 2000 ml, resolution 10 mlAdult range:
Tidal volume setting range
0.3 - 20 l/minInfant range:
0.5 - 60 l/minAdult range:
Minute volume setting range
± 15% or ± 2 cmH
2
O, whichever is
greater
0 - 120 cmH
2
OInspiratory pressure
(Pressure controlled modes)
0 - 80 cmH
2
O, resolution 1 cmH
2
OInfant range:
0 - 120 cmH
2
O, resolution 1 cmH
2
OAdult range:
Inspiratory pressure
setting range
Ye sCompressible volume
compensation
200 l/min (3.3 l/s)Max. inspiratory flow
4 – 100 ±1 breaths/minuteBreathing frequency
1:10 - 4:1I:E (VC, PC)
0 – 50 cmH
2
OPEEP
Flow / PressureTrigger
0 to 30% or 0 - 1.5 sInspiratory pause (VC)
276
FLOW-i 4.2, User's Manual
| 13
| Technical specifications |