3.7.1 Power supply and fuses
1
2
3
4
5
6
7
10
9
8
1. Auxiliary power outlets fuse 2A (option)
2. Auxiliary power outlets (option)
3. Auxiliary power outlets fuse 1A (option)
4. Auxiliary power outlets fuse 1A (option)
5. Patient monitor fuse (option)
6. Isolation transformer fuse (option)
7. Mains power inlet fuses
8. Lift fuses (C30 model only)
9. Mains power inlet
10. Equipotential terminal (earth)
Auxiliary power supply outlet connections vary
depending on country specific standards.
In addition to the individual fuses, the
maximum current delivered through the
system is regulated by a shared fuse.
In case of mains power failure, a battery
backup will power the system for a limited
time.
The patient monitor outlet and the auxiliary
power outlets do not have a battery backup.
28
FLOW-i 4.2, User's Manual
| 3
| System overview |