WARNING!
Restrictions apply to the use of auxiliary
power outlets:
• Outlets must not be used to power life
support equipment unless the life support
equipment itself is equipped with battery
backup.
• A multiple socket extension cord must
not be connected to any of the outlets.
CAUTION: The auxiliary power outlets shall
only be used for supplying power to
equipment intended to form part of the
medical electric system.
For more details on the power supply and
status, see Chapter 10, page 203.
3.7.2 Gas connections
2
1
5
6
3
4
7
1. O
2
outlet (option)
2. O
2
inlet
3. Air outlet (option)
4. Air inlet
5. N
2
O inlet
6. AGS outlet (country-specific connections)
7. Gas cylinder inlets for O
2
, Air and N
2
O
(option)
If equipment with high gas consumption are
connected to the gas outlets, the central gas
supply pressure should be above 3 bar (300
kPa, 44 PSI). Make sure that the central gas
supply is sufficient for the extra equipment.
FLOW-i 4.2, User's Manual