The intended use for the Auxiliary O
2
outlet is
to provide oxygen for patient therapy.
The intended use for the Auxiliary AIR outlet
is to provide drive gas for small ejector driven
suction devices.
Maximum allowed air flow from each of the
O
2
and Air outlets is 60 l/min.
The central N
2
O gas supply (wall supply) is
connected to the system via a pressure
regulator attached to the N
2
O gas inlet.
Clinical situations that require high gas
consumption might affect the supply gas
pressure.
If the gas supply is low, i.e. close to 2.5 bar
(250 kPa, 36 PSI), the system's ability to
deliver gas according to all possible settings
might be compromised by additional gas
hoses and connections.
3.7.3 Input/Output ports
1
2 3 4
5
6
1. RS232 serial data communication ports
2. Control panel connection
3. VGA connector
4. USB port
5. Network connection
6. Cable restrainer
The intended use of the USB port is only for
USB memory flash drives. Restricted items
include, but are not limited to, external hard
drives, USB hubs and any equipment using
the USB port primarily as a power source.
30
FLOW-i 4.2, User's Manual
| 3
| System overview |