3.10.3 Storage
During storage keep the system connected
to the mains power supply so that the
batteries maintain a full charge.
If the system is disconnected from a power
source, ensure the batteries are fully
charged before storage to avoid
deterioration of battery performance.
When the system is disconnected from a
mains power supply, a fully charged battery
can be stored in the FLOW-i system for up
to six weeks at temperatures between +5°C
(+40°F) and +40°C (+105°F). At
temperatures between +50°C (+125°F) and
60°C (+140°F) storage time is one week. If
these limits are exceeded, battery
performance can no longer can be
guaranteed.
MAQUET recommends that the vaporizers
are emptied before long term storage (>30
days).
For fire safety purposes, the system
requires functional batteries for system
Startup.
Ensure the system is not exposed to
temperatures below -25
o
C (-13
o
F) or above
+60
o
C (140
o
F).
For information on CO
2
absorber storage,
see Chapter 7, page 128.
Ensure the system is not exposed to a
relative humidity above 95%.
FLOW-i 4.2, User's Manual
39
| System overview |
3 |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals