Additional descriptionPreparatory steps
A
Step 1
Check that the breathing circuit is correctly
mounted and connected to the test plug (A).
Step 2
Check the water trap and sampling line.
Discard/empty any water present in the water trap.
Step 3
Check that the absorber is correctly mounted and
unsaturated. Also check that the switch is in the
'locked' position.
52
FLOW-i 4.2, User's Manual
| 4
| Startup and system checkout |