Additional descriptionPreparatory steps
max
min
max
min
Step 4
Check that the vaporizer(s) contains sufficient
agent.
Step 5
Check that the Anesthesia Gas Scavenging (AGS)
flow indicator is above the dashed area.
CAUTION: Ensure that the AGS is correctly
connected. Running the system with a poor AGS
connection may result in anesthetic agent being
emitted into the operation environment.
Step 6
Check that the central gas supplies are connected.
FLOW-i 4.2, User's Manual
53