Additional descriptionPreparatory steps
Step 7 (only if backup cylinders are installed)
Check that the Backup cylinders are opened. Close
the cylinders after completing System checkout.
The system checkout will not initiate if the system
is not connected to the central gas supply.
WARNING! If the backup gas supply pressure
is higher than the central gas supply pressure,
the backup cylinder gas will be used during the
System checkout. Make a note of the backup
cylinder pressure presented on the Control Panel
screen during the System checkout.
Step 8
Check that the manual resuscitator is readily
available and is working correctly.
Step 9
Check for adequate suction pressure in the suction
unit and that the auxiliary O
2
flow is functioning
properly, refer to Chapter 7, page 129, for
instructions.