Tidal/Minute volume
312
01-01 13 00
Sets the tidal/minute volume.
Different setting intervals apply for adult and
infant patient categories, see table below.
Minute
Volume
Tidal Volume
Patient
Category
0.5 - 60 l/min100 - 2000 mlAdult
0.3 - 20 l/min20 - 350 mlInfant
Either tidal (MTi) or minute (MV) volume can
be used and is selected as part of the system
configuration, see Chapter 9.
Pressure control above PEEP
312
01-01 13 00
Sets the pressure level above PEEP. When
the set inspiration pressure above PEEP is
reached, the system generates a pressure
plateau until the start of the expiration phase.
Different setting intervals apply for adult and
infant patient categories, see table below.
PC above PEEPPatient category
0 - 120 cmH
2
OAdult
0 - 80 cmH
2
OInfant
96
FLOW-i 4.2, User's Manual
| 5
| System functionality |