Cleaning / Disinfection / Sterilisation
Cleaning and sterilising Maquet Sterigrip sterilisable handles
6
Maquet PowerLED II
IFU 01811 EN 10
101 / 116
6.2 Cleaning and sterilising Maquet Sterigrip sterilisable handles
6.2.1 Preparation for cleaning
To prevent any soiling from drying out, soak the handles in a detergent-disinfectant bath contain-
ing no aldehydes, immediately after use.
6.2.2 Manual cleaning
1. Immerse the handles in a detergent solution
2
for 15 minutes.
2. Wash using a soft brush and a lint-free cloth.
3. Check that the handles are perfectly clean, with no remaining soiling. If not, use an ultra-
sound cleaning process.
4. Rinse thoroughly with clean water to fully eliminate the detergent solution.
5. Leave to air dry or wipe the handle with a dry cloth.
6.2.3 Cleaning in a washer-disinfector
Handles may be cleaned in a washer-disinfector and rinsed at a maximum temperature of 93°C.
Typical recommended cycles:
Step Temperature Time
Pre-wash 18-35°C 60 sec
Wash 46-50°C 5 min
Neutralisation 41-43°C 30 sec
Wash 2 24-28°C 30 sec
Rinse 92-93°C 10 min
Dry air dry 20 min
Tab.32: Typical cleaning cycles in a washer-disinfector
2
The use of non-enzymatic detergents is recommended. Enzymatic detergents may damage
the handles. Never soak the handles in these detergents for prolonged periods. Rinse thor-
oughly.

Table of Contents

Other manuals for Getinge Maquet PowerLED II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet PowerLED II and is the answer not in the manual?

Getinge Maquet PowerLED II Specifications

General IconGeneral
BrandGetinge
ModelMaquet PowerLED II
CategoryMedical Equipment
LanguageEnglish

Related product manuals