Technical specifications
EMC declaration
114 / 116
Maquet PowerLED II
IFU 01811 EN 10
Test type Test method Range of frequencies Boundaries
Measurement of conduc-
ted emissions on the
main ports
EN 55011 GR1 CL
A
7
0.15/0.5 MHz 79 dBμV QP
66 dBµV A
0.5/5 MHz 73 dBμV QP
60 dBµV A
5/30 MHz 73 dBμV QP
60 dBµV A
Measurement of the radi-
ated electromagnetic field
EN 55011 GR1 CL
A
7
30/230 MHz 40 dBµV/m PQ 10m
230/1000 MHz 47 dBµV/m PQ 10m
Tab.50: EMC declaration
Test type Test method Test level: healthcare facility
Electrostatic discharge im-
munity
EN 61000-4-2 Contact: ±8kV
Air: ±2; 4; 8; 15kV
Immunity to radiated RF elec-
tromagnetic fields
EN 61000-4-3
80 MHz, 2.7 GHz
3V/m Mod AM 80%/1kHz
Wireless RF frequencies
9 to 28V/m Mod AM 80%/1kHz
Immunity to fast electrical tran-
sients and bursts
EN 61000-4-4 AC: ±2kV - 100kHz
IO >3m: ±1kV - 100kHz
Immunity to power source
voltage surges
EN 61000-4-5 ±0.5; 1 kV diff.
±0.5kV, ±1kV, ±2kV common mode
Immunity to conducted inter-
ference due to electromagnetic
fields
EN 61000-4-6 150 kHz, 80 MHz
3Vrms Mod AM 80%/1kHz
ISM
6Vrms Mod AM 80%/1kHz
Immunity to voltage dips and
short interruptions
EN 61000-4-11
0% Ut, 10ms
(0°; 45°; 90°; 135°; 180°; 225°; 270°; 315°)
0% Ut, 20ms
70% Ut, 500ms
0% Ut, 5s
Harmonic current emissions EN 61000-3-2 Class A
Voltage variations, voltage fluctu-
ations, and flicker in public low-voltage
power supply networks
EN 61000-3-3 Compliant
Tab.51: EMC declaration
8.6.1 FCC Part 15 (USA only)
This equipment has been tested and found to comply with the limits for a ClassA digital device, pursuant to
Part15 of the FCC rules. These limits are designed to provide reasonable protection against harmful inter-
ference when the equipment is operated in a commercial environment. This equipment generates, uses,
and can radiate radio frequency energy and, if not installed and used in accordance with the instructions,
may cause harmful interference to radio communications. Operation of this equipment in a residential area
is likely to cause harmful interference, in which case the user will be required to suppress the interference at
its own expense.
7
The emission characteristics of this device enable it to be used in industrial areas and hos-
pital settings (Class A as defined in CISPR 11). If used in a residential environment (for which
class B defined in CISPR 11 is normally required), this device may not provide sufficient pro-
tection for radio frequency communication services. The user may need to take corrective
measures, such as relocating or re-orienting the device.

Table of Contents

Other manuals for Getinge Maquet PowerLED II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet PowerLED II and is the answer not in the manual?

Getinge Maquet PowerLED II Specifications

General IconGeneral
BrandGetinge
ModelMaquet PowerLED II
CategoryMedical Equipment
LanguageEnglish

Related product manuals