4.7 Positioning the screen holder(s)
4.7.1 Handling and positioning the screen holder
WARNING!
Risk of infection
The sterilisable handle is the only sterilisable component of the device. The
monitor, the screen holder and its accessories are not sterile and any contact
with the sterile team results in a risk of infection for the patient.
During the operation, the screen, the screen holder and its accessories must
never be touched by the sterile team and the handle must never be touched
by non-sterile personnel.
WARNING!
Risk of infection or tissue reaction
A collision between the device and another item of equipment may result in
particles falling onto the surgical site.
Pre-position the device before the patient arrives. Move the device carefully
to avoid a collision.
WARNING!
Risk of injury
A wrong handling of XHD1 screen holder may result in a hand injury.
Respect safety indications on the product.
Handling of the screen holder by the sterile team
Fig.105: Handling by sterile team
1. Move the device by grasping the sterilis-
able handle 1 or the DEVON or
DEROYAL sterile handle.
Handling of the screen holder by the non-sterile team
Fig.106: Handling by the non-sterile team
1. Move the device by grasping the flat-panel
monitor 2 , the screen holder frame 3 ,
the fork handle 4 or the rear box 5 .