Safety-related information
Safety instructions
2
Maquet PowerLED II
IFU 01811 EN 10
35 / 116
WARNING!
Risk of injury
If the battery discharges too quickly, a lighthead may go out during a proced-
ure.
Perform a battery lifetime test monthly to estimate the battery lifetime. Con-
tact the Getinge technical department if a malfunction occurs.
WARNING!
Risk of burns
This device is not explosion-proof. Sparks, which would not normally be haz-
ardous, may cause fires in oxygen-enriched atmospheres.
Do not use the device in environments rich in flammable gases or oxygen.
WARNING!
Risk of injury/infection
The use of a damaged device may lead to a risk of injury for users or a risk of
infection for patients.
Do not use a damaged device.
2.2.2 Electrical
WARNING!
Risk of electric shock
Anyone not trained in installation, maintenance or decommissioning opera-
tions is exposed to the risk of injury or electric shock.
Installation, maintenance and decommissioning of the device or components
of the device must be performed by a Getinge technician or a Getinge-trained
service technician.
WARNING!
Risk of injury
If a power cut occurs in the middle of an operation, the lightheads will go out
if the lighting system does not have a backup supply.
The hospital must comply with applicable standards on premises for medical
use and must have a backup power system.
2.2.3 Optical
WARNING!
Risk of burns
The high intensity of the light source results in a risk of burns to the eye if
the user looks directly towards the lighthead.
The patient’s eyes must be protected during facial surgery. Users must not
look directly into the light source.