Backup battery
94 / 116
Maquet PowerLED II
IFU 01811 EN 10
4.10.2 Performing battery tests
WARNING!
Risk of injury
A battery lifetime test fully discharges the batteries.
Do not perform an operation immediately after a battery lifetime test. Allow
time for the batteries to charge.
4.10.2.1 From the wall-mounted control keypad
Fig.120: Battery tests from the wall-mounted keypad
Running a battery backup test
1. Turn off the light.
2. Press Battery backup test 1 .
Ø If the test is successful, the battery level indicator 2 flashes green. If the test fails, the
battery level indicator 2 flashes red.
3. If the test fails, contact the Getinge technical service department.
4. Press Battery backup test 1 again.
Ø The battery level indicator 2 stops flashing. The light is on and ready for use.
Running a battery lifetime test (only with a Getinge backup)
1. Turn off the light.
2. Press Battery lifetime test 3 .
Ø If the test is successful, the battery level indicator 2 flashes green. If the test fails, the
battery level indicator 2 flashes red.
3. If the test fails, contact the Getinge technical service department.
Ø The light turns off when the test is complete.
4. Press Battery lifetime test 3 again.
Ø The battery level indicator 2 stops flashing.
NOTE
The battery lifetime test can be stopped at any time by pressing Battery lifetime test 3
again until the lightheads turn off.